Physical Dimensions:

  • Size:6 lbs
  • Dimensions:10x8x4 (wxhxd)

Safety Approval & Quality System:

  • Designed to meet IEC60601-1-1988, EN60601-1-1, EN60601-2
  • Class II Equipment, double insulated
  • Type BF applied parts
  • ISO9001 & EN46001 Certified

Power Requirements:

  • Power Supply: AC 90-264V/47-63Hz
  • Input Power: ≤55VA
  • Fuses: Two fuse sockets in the rear panel indicated by “FUSE”, OI 5X20, 2A/250V
  • Battery: 12V/4.0AH sealed lead-acid
  • Charge time: ≥4 hours
  • Operating time: ≥2 hours (full recharge)
  • Battery Charging Method: Automatic charging after monitor is connected to AC power supply (with charge protection function)
  • Discharge Protection: When powered by battery, the monitor will be automatically turned off when battery power is almost used up

ECG:

  • Patient Safety Standard: IEC60601-1-1988
  • CMRR ≥60dB (Common Mode Rejection Ratio)
  • Heart Rate Range: 20 ~ 254bpm 1bpm
  • Heart Rate Averaging: 8 second average
  • ST Segment Range: -0.8 ~ + 0.8mV
  • Interface: AAMI 6-pin
  • Lead Selection: I, II, III (3 lead mode)
  • I, II, III, aVR, aVL, aVF, V (5 lead mode) (ST and Arrhythmia analysis)
  • Lead Fault Alarm: Audible, Visual Input 5-lead ECG patient cable
  • QRS Indicator: Audible and Visual Alert
  • Waveform Storage: 6 minutes
  • Sweep Speed: 12.5/25/50 mm/sec
  • Gain Selection:4mV, 2mV, 1mV, 0.5mV, 0.25mV,Auto
  • Trends: 2 hours -> 4 hours -> 8 hours -> 24 hours -> 48 hours
  • Patient Isolation Breakdown: voltage 4000VAC 50Hz 60 seconds Leakage current <10μA
  • Frequency width Monitoring: mode 0.5 ~ 40Hz (+0.4dB,-3.0dB) Surgery mode 0.5 ~ 20Hz (+0.4dB, -3.0dB) not calibration significant
  • Patient Drive Current: <10μA
  • Enclosure Leakage Current: <0.1mA
  • Maximum T Wave Rejection Capability: 1.2mV
  • Heart Rate Alarm Response Time:< 7 seconds
  • Aspect Ratio: 0.24 ~ 0.6 sec/mV
  • Alarm Frequency Low alarm: 2-2.4kHz High alarm: 3-3.4kHz
  • Defibrillator Protected & ESIS Protected Tested with 5kV
  • Recovery Time Following Defibrillation: <5 seconds

RESP:

  • Measurement Method: Thoracic Impedance
  • Respiration Rate Range: 0 ~ 1001rpm
  • Accuracy: 2 rpm

Masimo SET SpO2

  • Measuring Range:
    • SpO2: 1?100 %
    • PR: 25?240 bpm
    • Perfusion: 0.02?20 %
  • SpO2 Accuracy (non-motion):
    • Adult Pediatric: 70?100 %: ±2 %, 0?69%: unspecified
    • Neonate: 70?100 %: ± 3 %,0?69 %: unspecified
  • SpO2 Accuracy (motion):
    • Adult Pediatric: 70?100 %: ±3 %, 0?69% : unspecified
    • Neonate: 70?100 %: ± 3 %, 0?69 % : unspecified
  • PR Accuracy (non-motion):
    • Adult Pediatric Neonate: 25?240 bpm: ±3 bpm
  • PR Accuracy (motion):
    • Adult Pediatric Neonate: 25?240 bpm: ±5 bpm
  • Low Perfusion Performance: >0.02 % Pulse Amplitude and % Transmission >5
    • SpO2: ±2 %
    • PR: ±3 bpm
  • Modes:
    • Averaging mode: 2,4,8,10,12,14 and 16 s
    • Sensitivity: Normal, APOD and Maximum

NIBP:

  • Method Automatic oscillometric
  • Parameters Systolic, diastolic, mean arterial pressure, pulse
  • Scale mmHg or kPa
  • Operating Modes Manual, Automatic, Continuous
  • Repeat Cycles 1 ~ 10, 15, 30, 60, 90, 120 minutes
  • Determination Systolic, Adult/pediatric 40 ~ 250mmHg (5.3 ~ 33.3kPa) Systolic, Neonate 20 ~ 160mmHg (2.7 ~ 21.3kPa) Diastolic, Adult/pediatric 10 ~ 180mmHg (1.3 ~ 24.0kPa) Diastolic, Neonate 10 ~ 140mmHg (1.3 ~ 18.7kPa)
  • Cuff Pressure Range Adult/pediatric 0 ~ 300mmHg (0 ~ 40.0kPa) Neonate 0 ~ 140mmHg (0 ~ 18.7kPa)
  • Initial Cuff Inflation Adult/pediatric 17010mmHg (22.7 1.3kPa) Neonate 10010mmHg (16.0 1.3kPa)
  • Deflation Pressure 30mmHg(4.0kPa) higher than the last systolic pressure
  • Cuff Inflation Rate No greater than 50mmHg/sec
  • Measurement Time Typical 25 seconds, Maximum 40 seconds, Typical Stat 20 seconds
  • Pressure Display Accuracy 3mmHg
  • BP Pulse Rate Accuracy 2% @ 40 ~240bpm
  • Cuff Neonate, infant, pediatric,standard adult

TEMP (Dual Channel)

  • Range 0 ~ 50C
  • Probe YSI 400 Skin surface or rectal /esophageal
  • Scale Celsius
  • Accuracy 0.1 C
  • Resolution 0.1 C

CO2 (Respironics) (mainstream or sidestream)

  • Type Side stream, mom-dispersive IR
  • CO2 Range 0-99mmHg
  • Scale mmHg/kPa
  • Accuracy +- 2mmHg (0-40mmHg) +- 5mmHg (41-76mmHg) +-10mmHg (77-99mmHg)
  • Calibration Automatic
  • Respiration Range 0-150rpm, +- 2rpm

TFT Color Display

  • Size 8 inches
  • Matrix 640 (H) x 480 (V) pixels
  • Recorder (Optional)
  • Type Built-in 2-channel thermal array recorder
  • Print mode Text or waveform
  • Waveforms Real time or alarm-triggered
  • Resolution 400dpi vertical, 800dpi horizontal
  • Annotations Time, date, vital sign readings